Shim Camera

The Passive Shim Camera is an essential tool for installing MRI systems that require passive shimming. By accurately measuring the magnetic field, this comprehensive kit allows engineers to add thin metal plates inside the passive shim tray to create magnetic field homogeneity. The kit includes a PT2026 teslameter, combined with dedicated Camera Software & Hardware, an MFC2046 Upgrade Conversion Kit, and a half-moon probe array, available for both 1.5T and 3.0T systems. The system is completed by user-friendly conversion software that streamlines the shimming process and a specialized shim tray puller for safe and efficient handling of shim trays. By employing the Passive Shim Camera, MRI engineers can achieve a flawless workflow.

No items found.
GE 1.5T Option
Box#
Part Number
Description
[1]
PT2026
PT2026 Teslameter with Camera Option Hardware/Software
[1]
MFC2046upgrade
MFC2046 Conversion Kit: Field Camera Amplifier, 10M Cable, Camera Switch
[1]
9046-32-1-500-1.5T-R-R
GE 1.5T halfmoon Probe array, Field: 63.86MHz
GE 3.0T Option
Box#
Part Number
Description
[1]
PT2026
PT2026 Teslameter with Camera Option Hardware/Software
[1]
MFC2046upgrade
MFC2046 Conversion Kit: Field Camera Amplifier, 10M Cable, Camera Switch
[1]
9046-32-1-500-3.0T-R-R
GE 3.0T halfmoon Probe array, Field: 127.72MHz
Philips 1.5T Option
Box#
Part Number
Description
[1]
PT2026
PT2026 Teslameter with Camera Option Hardware/Software
[1]
MFC2046upgrade
MFC2046 Conversion Kit: Field Camera Amplifier, 10M Cable, Camera Switch
[1]
Philips-1.5T-HMPA
Philips 1.5T halfmoon Probe array, Field: 63.86MHz
Philips 3.0T Option
Box#
Part Number
Description
[1]
PT2026
PT2026 Teslameter with Camera Option Hardware/Software
[1]
MFC2046upgrade
MFC2046 Conversion Kit: Field Camera Amplifier, 10M Cable, Camera Switch
[1]
Philips-3.0T-HMPA
Philips 3.0T halfmoon Probe array, Field: 127.72MHz
PT2026
Features
  • Ultra-high resolution less than 10 ppb at 3 T
  • Accuracy of 1 ppm
  • Measurement range from 38 mT to 30 T
  • Integrated Hall sensor
  • Ultra-thin probes of 1.5 mm thickness
  • Gradient tolerance > 1,500 ppm/cm at 1 T
  • Acquisition speed of 33 Hz
  • FID and FFT analysis
  • ISO17025 calibration
Frequency Range
1 MHz - 1.1GHz
Resolution
  • ± 0.1 Hz (stable field, low gradient, no averaging)
  • Less than 0.01 ppm (10 ppb) in uniform 3 T field
Accuracy
± 5 ppm, independent of temperature
Max gradient
> 1000 ppm/cm
Measurement rate
Up to 33 Hz
Trigger modes
Immediate, Timed, Bus, External
Power
55 VA, 100 – 240 VAC, 50-60 Hz
Overvoltage
Accepts temporary overvoltage occurring on the mains supply transient overvoltage up to overvoltage category ll levels
Fuse
3.15 A (T), 5x20 mm, 250 V
Environment
Indoor use; no air inlet (IP 50)
Operating temperature
10 – 40 °C
Storage / transport temperature
-25 – 80 °C
Altitude
Maximum 80 % for temperatures up to 31 °C, decreasing linearly to 50% relative humidity at 40 °C
Magnetic environment
Less than 0.2 T
MFC2026Upgrade
Features
  • Measuring fields up to 1.1 GHz or 30 T
  • Large variety of probe array geometries
  • Up to 255 sensors
  • Enhanced flexibility
  • Wide-range probe for field ramping
  • User-friendly software
Environment
Indoor use; IP 50
Operating temperature
10 – 40 °C
Altitude
≤ 2000 m
Relative humidity
Maximum 80 % for temperatures up to 31 °C, decreasing linearly to 50% relative humidity at 40 °C
Magnetic environment
Less than 1 T
GE 1.5T Halfmoon Probe Array
Features
  • Tailored to the dimensions, geometries, and magnetic field strengths of your MRI systems
  • For DSV from 100 mm to 600 mm
  • Standard configurations with 24 or 32 probes
  • Option to integrate a wide-range central probe
  • Factory normalized to ±0.2 ppm
Frequency
  • 1 MHz - 1.1 GHz
  • Depends on the magnetic field requested, probe tuning dedicated to one frequency
Magnetic Field Range Probes on the periphery of the half-moon
  • Nominal probe-array value ±3% (typical)
  • 1.5 T probe-array: range => 1.455 T to 1.545 T
  • 3.0 T probe-array: range => 2.91 T to 3.09 T
Probe-array normalization
The discrepancy between probes placed in the exact same field ≤ ±0.2 ppm
Magnetic Field Range Central wide-range probe
  • Dynamic range of x3 below the nominal probe-array value
  • 1.5 T probe-array: wide-range probe => 0.5 T to 1.5 T
  • 3.0 T probe-array: wide-range probe => 1.0 T to 3.0 T
Cable length
4m
Dimensions
337.4 x 29.4 x 524.4 mm
[GE]

  ☑ Signa 3.0T Excite (WB/LCC300)

  ☑ Signa 3.0T HD (WB/LCC300)

  ☑ Signa 3.0T HDxt (WB/LCC300)

  ☑ Optima MR450W 1.5T (HM)

  ☑ Discovery MR750 3.0T (WB/LCC300)

  ☑ Discovery MR750W 3.0T (UA)

  ☑ Signa Premier 3.0T (WB/LCC300)

  ☑ Signa 1.5T Artist (PM)

  ☑ Signa 1.5T Prime (PM)

  ☑ Signa 1.5T Voyager (PM)

[Philips]

  ☑ Intera 1.5T (MR INT Magnet 15T)

  ☑ Intera 1.5T Achieva Pulsar (MR ACH Magnet 15T)

  ☑ Intera 1.5T Achieva Nova (MR ACH Magnet 15T)

  ☑ Intera 1.5T Achieva Nova Dual (MR ACH Magnet 15T)

  ☑ Intera 1.5T Achieva CV Nova (MR ACH Magnet 15T)

  ☑ Achieva 1.5T (MR ACH Magnet 15T_M)

  ☑ Intera 3.0T Achieva Quasar Dual (MR ACH Magnet 30T F2000)

  ☑ Intera 3.0T Achieva  Achieva XR (MR ACH Magnet 30T F2000)

  ☑ Achieva 3.0T (MR ACH Magnet 30T_M REX)

  ☑ Achieva 3.0T TX (MR ACH Magnet 30T_M REX)

  ☑ Ingenia 1.5T (MR WA Magnet15)

  ☑ Ingenia 3.0T (MR WA Magnet30)

  ☑ Multiva 1.5T (MR Magnet15T_1 F2000)

User Manual